Techincal specifications
System:
- Fully automated,random access analyzer for clinical Chemistry, Electrolytes & Homogeneous Immunoassays(HIA).
Test throughput:
- Up to 450 tests/hr.
Reagent concept:
- Barcoded cobas c reagent packs with convenient packaging & ready –to –use reagents.
Onboard reagent capacity:
- Up to 45 assays (42 cooled cassette slots + channels for ISE).
Programmable parameters:
- Up to 117 photometric, 3 Ise tests, 8 calculated parameters & 3 serum indices.
Sample types:
- Serum, Plasma, Urine, CSF & Whole Blood.
Onboard sample capacity:
- 108 positions with continuous access & STAT interrupt capability.
Sample container types:
- Primary tubes: 5-10ml (16x 100, 16x75, 13x100, 13x75 mm)
- Sample cup : 2.5 ml
- Micro cup : 1.5 ml
- Cup on tube
Sample volume:
- to 35 µl in 0.1 µl steps.
Sample clot detection:
- Yes
Minimum sample volume:
- Primary tubes: 700 µl
- Sample cup : 100 µl
- Micro cup : 50 µl
Sample barcode types:
- Code 128, codabar (NW 7), Interleaved 2 of 5, code 39.
System interfaces:
- Touchscreen PC, RS 232 serial interface, bi-directional.
- Cobas link (TeleService and auto download).
Sample database:
- 10,000 routine/STAT samples.
Test methods:
- 1 point, 1point+ prozone check, 2 point, 2 point kinetic, 2 point + prozone check,3 point, 1 point + kinetics, rate a, rate a + serum index, rate a with blank, Rate B.
Calibration methods:
- Linear, non-linear multi-points, 2 point calibration, K-factor, up to 100 calibrators pre-programmable, storage of up to 180 curves; Preventive calibration of stand – by cassettes.
QC methods:
- Real-time, individual & cumulative QC; Auto QC; Up to 100 different controls pre-pro-grammable, Preventive QC after calibration of stand-by cassettes.
Rerun/repeat/reflex:
- Automatic rerun/manual reruns, repeats & automatic reflex supported.
Electrical requirements:
- Power requirements:230/400 Volts AC 3/N/PE, 110 Volts AC 1.5 kVA.
- Frequency : 50Hz or 60hz +/- 0.5%.
Water/waste requirements:
- Water quality :Bacteria free, de-ionised water supply resistnce of 1.0 µS/cm max.
- Water consumption: Up to 12 l/hr during operation.
- Water Pressure : 0.5- 3.5 kg/cm3 (49- 343 KPa).
- Water temperature :15 to 30®C.
- Biohazards waste : Separate container located in the system.
- Wall drain for waste: 100 mm above the floor.
Regulatory compliance:
- GS,CE, UL, C-UL compliant.
Operating conditions:
- Ambient temperature/humidity: 15 to 32®C/30 to 85% (RH, without condensation)
·